Skip to Content

Radiated Emissions and Radiated Immunity EMC Testing Services

TEF-Health Service
Physical
Pricing/Discount Options: Call #2
Service Image

Service Description

Overview

This service evaluates the radiated electromagnetic compatibility of industrial, scientific, medical and multimedia electronic equipment. Radiated emission testing measures unwanted electromagnetic energy generated by the Device Under Test (DUT), while radiated immunity testing evaluates its ability to operate correctly when exposed to external electromagnetic fields.

The service supports product development, pre-compliance, troubleshooting and regulatory conformity. Testing is performed according to applicable EMC standards, including CISPR 11, CISPR 32, IEC 61000 series and IEC 60601-1-2 for medical electrical equipment.

How can the service help you?

The service helps manufacturers identify EMC issues before product qualification and market introduction. It can be used to:

  • verify compliance with radiated emission limits;
  • assess immunity to external RF fields;
  • identify critical frequencies and EMC weaknesses;
  • support troubleshooting and design improvements;
  • provide documented results for product qualification.

How will the service be delivered?

Testing is performed in Multitel's anechoic chamber in Mons, Belgium. The facility can accommodate both conventional electronic equipment and large industrial systems, with DUTs up to approximately 1.5 m in diameter and 1 ton in weight. The test configuration, applicable standards, operating modes and acceptance criteria are defined with the customer before testing. This is done during a first onsite or online meeting which can be organised within days after the matchmaking. Once this is done, the tests in the anechoic chamber will start typically 2-3 weeks after this meeting depending on the availability of the chamber. We always plan for sufficient time so that all the necessary tests can be conducted and the company does not need to plan for another slot.

Keywords: EMC electromagnetic emission standards radiated
Offerings: Hardware (Design, Development, Manufacturing, etc.) Testing Conformity & Compliance Hardware (CE, MDR, usability, etc.) Testing Pre-Clinical (Neurology, Cardiology, wearables, motion, psychophysical, GLP, animal safety, toxicity, etc.)
TEF-Health Use Case Domain: all
Provider Logo

Provider & Contact

Provider Organisation Multitel (MULTITEL)
Provider Country Belgium
Organisation Website https://www.multitel.be/
Billing: per day
Full Price 1400–2000 EUR
Reduced Price 1100-1700EUR
Pricing Detail
  • Full price: 1400-1700 EUR per day First day with setup installation & Engineer support: 1700€ per day Other days - Engineer support: 1400€ per day
  • Reduced price: 1100-1300 EUR per day First day with setup installation & Engineer support: 1300€ per day Other days - Engineer support: 1100€ per day

Operational Details

Service Inputs Medical device to be tested
Service Outputs EMC pre-test results
Dependencies & Restrictions Facility costs;MDR
Certification Support Service outputs can be used to obtain the CE marking regarding EMC testing.
  • Service outputs can be used to obtain the CE marking regarding EMC testing
Service Standards
  • CISPR11/CISPR32/EN 61000-4-3